C C A A U U T T I I O O N N
• Do not use high frequency surgical instruments or endocardial catheters while the A A l l t t r r i i x x system is in use. This is to
avoid the risk of electrical shock, burns, or electromagnetic interference.
Use only Stryker temperature probes. See Patient temperature probes (page 47).
To connect the temperature probe:
1. Inspect the temperature probe and reusable adapter cable for wear, breakage, or fraying. Replace, if necessary.
2. Align the red dot on the R R e e u u s s a a b b l l e e A A d d a a p p t t e e r r C C a a b b l l e e (B) to the controller (A) with the red dot on the patient probe port A or
port B.
3. Connect the plug (C) to the patient temperature probe.
4. Apply the temperature probe to the patient. Follow your hospital protocol and the manufacturer's directions for the
selected temperature probe use.
5. Tap Confirm, if applicable.
N N o o t t e e - - Temperature readings may vary between temperature measurement sites.
C C o o n n n n e e c c t t i i n n g g t t h h e e r r e e u u s s a a b b l l e e p p a a t t i i e e n n t t t t e e m m p p e e r r a a t t u u r r e e o o u u t t p p u u t t c c a a b b l l e e
This feature allows the operator to view the temperature on the A A l l t t r r i i x x system and on an external device. Always connect
the reusable patient temperature output cable to a 400 series compatible external device for temperature accuracy.
W W A A R R N N I I N N G G - - Always use Stryker accessories. Only IEC 60601-1 equipment shall be hooked to the patient temperature
ports. Failure to comply with these instructions may invalidate any or all warranties and may negatively affect the products
EMC performance. This also protects the product from cardiac defibrillation.
To connect the reusable patient temperature output cable:
1. Insert the reusable patient temperature output cable into the patient temperature port (Figure 8).
2. Connect the other end of the reusable patient temperature output cable to the external device.
N N o o t t e e
• When A A l l t t r r i i x x is powered, the calibration of the patient temperature output is completed.
• If you need to calibrate the patient temperature output cable, power cycle the product by removing the plug from the
wall.
EN
22
C
F F i i g g u u r r e e 7 7 – – P P o o r r t t s s e e l l e e c c t t e e d d
F F i i g g u u r r e e 8 8 – – P P a a t t i i e e n n t t t t e e m m p p e e r r a a t t u u r r e e o o u u t t p p u u t t p p o o r r t t
A
B
8001-109-005 Rev C.2