N N o o t t e e - - The term "thermal transfer devices" is used throughout this manual and is interchangeable with blankets and wraps,
unless indicated otherwise.
P P o o w w e e r r i i n n g g o o n n t t h h e e p p r r o o d d u u c c t t
The operator should stand in front of the controller within arm's reach. This allows the operator to see and respond to the
display notifications.
W W A A R R N N I I N N G G
• Avoid reduction in water flow. Do not connect two or more thermal transfer devices in a series on a single port.
• When you operate the product near ambient temperature limitations of 15.0° C (59.0° F) or 32.0° C (89.6° F), you may
experience a reduction in product performance.
C C A A U U T T I I O O N N
• Shock Hazard - Improper handling of the power cord may damage the power cord and cause potential shock hazards. If
damage has occurred to the power cord, immediately remove the temperature management system from service to
avoid the risk of serious injury or death. Contact the appropriate maintenance personnel.
• Electric shock. This equipment must only be connected to a supply mains with protective earth.
• Always plug this product directly into a properly grounded hospital-grade or medical-grade wall outlet to achieve
grounding reliability.
• Do not use high frequency surgical instruments or endocardial catheters while the A A l l t t r r i i x x system is in use. This is to
avoid the risk of electrical shock, burns, or electromagnetic interference.
• Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal or mask type.
• Do not place cables, hoses, or power cord in walkways to avoid the risk of trip hazards.
• Do not use three or more adult Mul-T-Blanket products at the same time to avoid the risk of water overflow when you
power off the controller.
To start the product:
1. Plug the power cord into a wall outlet.
2. Tap the Stand-by button to start the product.
3. If you are going into Automatic mode or Monitor mode, see Selecting and setting the primary probe (page 27). If manual
mode, got to the next step.
4. See Removing and replacing the reservoir (page 26).
5. See Filling the reservoir with sterile distilled water (page 26).
6. Connect up to three thermal transfer devices (with the exception of adult Mul-T-Blankets) to dedicated adapter hoses
and ports.
7. Open the clamps on the connector hose and the thermal transfer devices to provide proper water flow.
8. See Filling a thermal transfer device (page 27).
9. See Selecting a therapy mode (page 28).
10. Make sure that the desired port configuration is maintained and that water is flowing through the thermal transfer
devices.
W W A A R R N N I I N N G G - - Always turn or re-position the patient over the duration of therapy, if possible, to reduce the risk of
pressure ulcers. Follow your hospital protocol.
8001-109-005 Rev C.2