• The controller determines W W a a r r m m i i n n g g or C C o o o o l l i i n n g g therapy based on the selected water target temperature and the
current water temperature.
• Do not place additional heat sources between the patient and thermal transfer device.
• After the patient target temperature is achieved, patient temperature is controlled within +/-0.3° C.
• If the patient temperature is not within 0.5° C of the current target temperature, the yellow patient icon will flash and
the patient temperature deviation alarm will sound. This occurs after the initial patient target temperature is achieved.
S S e e t t t t i i n n g g o o r r e e d d i i t t i i n n g g t t h h e e c c o o o o l l i i n n g g r r a a t t e e s s
Setting cooling rates is for Automatic Mode only.
1. To set the cooling temperature, highlight your choice of cooling rates.
Select a cooling rate
Max
2. Tap Confirm.
3. Tap the Edit button to make changes.
S S e e t t t t i i n n g g o o r r e e d d i i t t i i n n g g t t h h e e w w a a r r m m i i n n g g r r a a t t e e s s
Setting warming rates is for Automatic Mode only.
1. To set the warming temperature, highlight your choice of warming rates.
Select a warming rate
Max
2. If you select Set Custom, tap the Increase and Decrease buttons to set the rate (Figure 18).
8001-109-005 Rev C.2
Description
Maximum: approaches patient target temperature as fast as possible
Medium: water is cooled to target, with a max of 15.0° C difference between the patient
and the water temperature
Minimum: water is cooled to target, with a max of 10.0° C difference between the patient
and the water temperature
Description
Maximum: approaches patient target temperature as fast as possible
Medium: patient temperature increases at a rate of 4.0° C in 12 hours (0.33° C/ hour).
Minimum: patient temperature increases at a rate of 4.0° C in 24 hours (0.17° C/hour).
Set Custom: patient temperature increases at a customized temperature and time
period the operator selects. The temperature increases 0.05° C/hour to 0.5° C/hour.
29
EN