E E M M C C I I n n f f o o r r m m a a t t i i o o n n
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ' ' s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c e e m m i i s s s s i i o o n n s s
The A A l l t t r r i i x x system is intended for use in the electromagnetic environment specified below. The customer or the user of
A A l l t t r r i i x x should assure that it is used in such an environment.
E E m m i i s s s s i i o o n n s s t t e e s s t t
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
N N o o t t e e - - The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment, for which CISPR 11 class B is normally required, this equipment might not
offer adequate protection to radio frequency communication services. The user might need to take mitigation measures,
such as relocating or reorienting the equipment.
R R e e c c o o m m m m e e n n d d e e d d s s e e p p a a r r a a t t i i o o n n s s d d i i s s t t a a n n c c e e s s b b e e t t w w e e e e n n p p o o r r t t a a b b l l e e a a n n d d m m o o b b i i l l e e R R F F c c o o m m m m u u n n i i c c a a t t i i o o n n s s e e q q u u i i p p m m e e n n t t a a n n d d t t h h e e
The A A l l t t r r i i x x system is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of A A l l t t r r i i x x can help prevent electromagnetic interferences by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the A A l l t t r r i i x x system as
recommended below, according to the maximum output power of the communications equipment.
R R a a t t e e d d m m a a x x i i m m u u m m o o u u t t p p u u t t
p p o o w w e e r r o o f f t t r r a a n n s s m m i i t t t t e e r r
W
0.01
0.1
1
EN
58
C C o o m m p p l l i i a a n n c c e e
Group 1
Class A
Class A
220-240V/50Hz
220V/60Hz
Does not apply to 100V
50/60Hz or 120V/60Hz
Complies
220-240V/50Hz only
A A l l t t r r i i x x s s y y s s t t e e m m
S S e e p p a a r r a a t t i i o o n n d d i i s s t t a a n n c c e e a a c c c c o o r r d d i i n n g g t t o o f f r r e e q q u u e e n n c c y y o o f f t t r r a a n n s s m m i i t t t t e e r r
1 1 5 5 0 0 k k H H z z t t o o 8 8 0 0 M M H H z z
D D = = ( ( 1 1 . . 2 2 ) ) ( ( √ √ P P ) )
0.12
0.38
1.2
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c e e n n v v i i r r o o n n m m e e n n t t
The A A l l t t r r i i x x system uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The A A l l t t r r i i x x system is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
m
8 8 0 0 M M H H z z t t o o 8 8 0 0 0 0 M M H H z z
D D = = ( ( 0 0 . . 3 3 5 5 ) ) ( ( √ √ P P ) )
0.035
0.11
0.35
8 8 0 0 0 0 M M H H z z t t o o 2 2 . . 7 7 G G H H z z
D D = = ( ( 0 0 . . 7 7 0 0 ) ) ( ( √ √ P P ) )
0.07
0.22
0.7