P P r r o o b b l l e e m m
Patient temperature
Patient temperature output
(PTO)
EN
50
P P o o s s s s i i b b l l e e c c a a u u s s e e
Out of range
External device output
displays high value >45° C
when input is out of range
(As a result of one of the
following: patient probe
disconnected, controller in
standby / sleep mode,
patient temperature outside
the range of 25° - 45°C).
A A c c t t i i o o n n
Check the placement of the
probe
To resume calibration:
•
Disconnect the external
device from the reusable
adapter cable
•
Tap the Help button to
display the alarm
screens
•
Find the "Temperature
Output Alarm" screen
•
Tap Confirm to restart
calibration
•
Wait until the "Monitor" is
solid
•
Connect the external
device to the reusable
adapter cable
R R e e m m o o v v e e f f r r o o m m u u s s e e
Not applicable
Not applicable
8001-109-005 Rev C.2