Electromagnetic Compatibility

Although this equipment conforms with the intent of the directive 89/336/EEC in relation
to electromagnetic compatibility (EMC), all electrical equipment may produce interference.
If interference is suspected, move equipment away from sensitive devices or contact the
manufacturer.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Radios, cell phones and similar devices may affect this equipment and should be kept at least
10.8 ft (3.3m) away from the equipment.
Medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information in the following tables.
The following tables document compliance levels and guidance from the IEC 60601-1-2;
2007 Standard, for the electromagnetic environment in which the AtmosAir MRS should be
used in a clinical environment.
Guidance and manufacturer's declaration - electromagnetic emissions
The AtmosAir MRS is intended for use in the electromagnetic environment specified below. The customer or user of the
AtmosAir MRS should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
16
Compliance
The AtmosAir MRS uses RF energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely to cause any interference in
Class B
Class B
yes
Electromagnetic environment
nearby electronic equipment.