Getinge Arjohuntleigh AtmosAir 9000A Instrucciones De Uso página 20

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 112
Guidance and manufacturer's declaration - electromagnetic immunity
The AtmosAir MRS is intended for use in the electromagnetic environment specified below. The customer or user of the
AtmosAir MRS should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
230VAC Pump
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50Hz/60Hz)magnetic field
IEC 61000-4-8
NOTE: U, is the a.c. mains voltage prior to application of the test level.
Recommended separation distances between portable and mobile RF communications
equipment and the AtmosAir MRS
The AtmosAir MRS is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the AtmosAir MRS can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the AtmosAir MRS as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
150 kHz to 80 MHz
not applicable
0.01
N/A
0.1
N/A
1
N/A
10
N/A
100
N/A
IEC 60601
Test Level
±6 kV Contact
±8 kV Air
±2 kV
1 kV line(s) to Line(s)
1 kV line(s) to Line(s)
2 kV line(s) to Earth
5% half cycle
40% for 5 cycle
70% for 25 cycle
5% for 5 seconds
3 A/M
Separation distance according to frequency of transmitter
meters
80 MHz to 800
MHz
3.5
P
d=[ — ]
E
1
0.12
0.37
1.2
3.7
12
Electromagnetic
Compliance Level
Environment Guidance
In accordance with IEC
60601-1-2:2007, floors
±6 kV Contact
are covered with synthetic
material, the relative
±8 kV Air
humidity should be at least
30%
±2 kV
N/A
N/A
2 kV line(s) to Earth
5% half cycle
40% 5 cycles
70% 25 cycles
N/A
5% for 5 seconds
Power frequency magnetic
fields should be at levels
characteristic of a typical
3 A/M
location in a typical
commercial or hospital
environment.
800 MHz to 2.5 GHz
7
P
d=[ — ]
E
1
0.23
0.74
2.3
7.4
23
17

Publicidad

Tabla de contenido
loading

Tabla de contenido