Descargar Imprimir esta página

Storz AUTOCON II 400 Manual De Servicio página 103

Publicidad

83
Technische Beschreibung
Technische Daten
POWER Cut
HF voltage form
Rated frequency
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
HF power limitation
Max. power output at
rated load resistance
TOP Cut
HF voltage form
Rated frequency
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
HF power limitation
Max. power output
at rated load resistance
LAP-C-Cut
HF voltage form
Rated frequency
Crest factor
Effect 1-4:
Effect 5+6:
Effect 7+8:
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
HF power limitation
Max. power output at
rated load resistance
Technical description
Technical data
unmodulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
1.4 (at RL = 500 ohms)
500 ohms
740 Vp
8
automatic control of HF peak voltage
10 W to 300 W in 1 W steps
300 W ± 20%
unmodulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
1.4 (at RL = 500 ohms)
500 ohms
950 Vp (with electric arc)
8
automatic control of arc intensity
10 W to 300 W in 1 W steps
300 W ± 20%
pulse-modulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
3.0 (at RL = 500 ohms)
3.2 (at RL = 500 ohms)
3.8 (at RL = 500 ohms)
500 ohms
1450 Vp
8
automatic control of HF peak voltage
10 W to 200 W in 1 W steps
200 W ± 20%
Descripción técnica
Datos técnicos

Hide quick links:

Publicidad

Capítulos

Solución de problemas

loading