Descargar Imprimir esta página

Storz AUTOCON II 400 Manual De Servicio página 106

Publicidad

86
Technische Beschreibung
Technische Daten
PAPILLO Cut
HF voltage form
Rated frequency
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
Max. power output
High frequency power limitation
GASTRO Cut
HF voltage form
Rated frequency
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
Max. power output
High frequency power limitation
BIPOLAR Cut
HF voltage form
Rated frequency
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
Max. power output
High frequency power limitation
C-Cut
HF voltage form
Rated frequency
Crest factor
Effect 1-4:
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
Max. power output
High frequency power limitation
Technical description
Technical data
unmodulated sinusoidal AC voltage
350 kHz ± 10%
1.4
75 ohms
800 Vp
4
automatic control of HF peak voltage
200 W ± 20%
10 W to 200 W in 1 W steps
unmodulated sinusoidal AC voltage
350 kHz ± 10%
1.4
75 ohms
880 Vp
4
automatic control of HF peak voltage
200 W ± 20%
10 W to 200 W in 1 W steps
unmodulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
1.4 (at RL = 500 ohms)
500 ohms
740 Vp
8
automatic control of HF peak voltage
100 W ± 20%
1 W to 200 W in 1 W steps
pulse-modulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
3.0 (at RL = 500 ohms)
500 ohms
1450 Vp
4
automatic control of HF peak voltage
200 W ± 20%
10 W to 200 W in 1 W steps