4.1 Tools required for the individual testing and
adjustment procedures
Note:
Prior to delivery the device is completely adjusted and tested by KARL STORZ. Re-adjustment
should only be carried out by fully-qualified technical staff authorized by KARL STORZ.
Note:
The results of the various tests should be recorded in a test report, see Chapter 7. Appendix.
The manufacturer's instructions detailed below for carrying out a safety inspection of the AUTOCON
400 are applicable to the following versions:
– 205352 20-111
– 205352 20-112
– 205352 20-113
– 205352 20-114
– 205352 20-115
– 205352 20-116
KARL STORZ recommends the following measuring and test equipment for conducting safe ty checks on
the AUTOCON
®
II 400:
– Bender Unimet 1100 safety tester
– Metron QA-ES MKII HF power meter
– Test adapter KARL STORZ-SCB
– AUTOCON
®
II 400 test equipment
– Digital multimeter
– Dual-pedal footswitch
4.2 Visual inspection
4.2.1 Exterior of the device
– General labeling:
Item number / device type
Potential equalization
– Adhesive label affixed and legible:
Model plate
CE mark
Serial number on the manufacturer's identification plate on the rear side
Adhesive fuse label on fuse drawer
Correctly rated fuses
– Power switch:
lights up "GREEN" when switched on
– Socket modules, HF socket rack and housing:
mechanically undamaged
– Display / touch screen:
not scratched
mechanically undamaged
4-2
– 205352 20U111
– 205352 20U112
– 205352 20U113
– 205352 20U114
– 205352 20U115
– 205352 20U116
®
bus termination
AUTOCON
MODEL 205352 20-1
Order no. W20686
Order no. SV4029
Order no. 200138 30
®
II 400
®
II
12.06 / V1.0