3.4 Troubleshooting

3.4.1 Error codes

Error code
Description of error
A-02
Communication error during activation
A-0A
NE module not available
A-0B
NE contact: resistance too low
A-0C
NE contact: resistance too high
A-0E
Erroneous ESI communication
A-0F
24 V supply voltage too low
Release activation (Termination by
A-10
autostop)
A-11
Missing activation signal
A-12
Activation during power up not allowed
A-17
Simultaneous activation not permitted
Missing deactivation signal from HF
A-18
module
A-19
Power fail signal detected
A-1C
Max. activation time exceeded
A-1E
Activated touch screen during power up
Critical NE symmetry (activation
A-1F
terminated)
Remove tissue contact, Autostop
A-22
reached
A-23
Erroneous message from HF module
A-25
Illegal socket module configuration
A-50
Incompatible ESI version
A-51
Incompatible Black Box version
Internal device temperature out of
A-52
specs (> 70 ˚C )
A-53
Low capacitance of back-up battery
Back-up battery empty (no data
A-55
storage)
A-56
Procedure level B not existent
A-60
Procedure transfer not possible (SCB)
A-81
Illegal ESI message
A-84
Dual-pedal footswitch recognized
A-85
Dual-pedal footswitch disconnected
A-88
Single-pedal footswitch recognized
A-89
Single-pedal footswitch disconnected
Missing status message from HF socket
A-8E
1
Missing status message from HF socket
A-8F
2
Missing status message from HF socket
A-90
3
A-93
Triple-pedal footswitch recognized
A-94
Triple-pedal footswitch disconnected
12.06 / V1.0
Solution
Check neutral electrode and reposition
if necessary
Check neutral electrode and reposition
if necessary
End activation of HF energy
Send instrument to service technician
Check footswitch, handpiece and cable
Activate either via footswitch or
handpiece
Check power cord, power outlet and
power supply
Cancel activation
Touch screen must not be pressed
during power-up
Check neutral electrode and reposition
if necessary
Break contact with tissue
Reconnect dual-pedal footswitch and
check cable
Reconnect single-pedal footswitch and
check cable
Reconnect triple-pedal footswitch and
check cable
AUTOCON
®
II 400
MODEL 205352 20-1
Priority
High
High
Medium
Medium
High
High
Low
High
Medium
Medium
High
Low
Low
Low
High
Low
High
Low
Medium
Medium
High
Low
Low
Low
Low
High
Low
Low
Low
Low
High
High
High
Low
Low
3-7

Hide quick links:

Publicidad

Capítulos

Solución de problemas

loading