Section 2 Installation
Brooks
®
GF40/GF80 Series
2-17 Using the MultiFlo
Figure 2-5 MultiFlo Cable Adaptor
2-10
TM
Configurator
The MutliFlo Configurator application is used to configure the gas and range
of the GF40/GF80 Series devices.
The MultiFlo Configurator application allows communication to GF40/GF80
Series devices through personal computer with serial COM Port and a
Windows XP or Windows 7 operating system. It's primary function is to
configure gas and flow ranges within defined standard configurations. Flow
ranges are configured to the Nitrogen equivalent.
Using the MultiFlo Configurator software, configure the gas and flow rate
according to Table 2-2.
Table 2-2 Gas and Flow Ranges - MultiFlo Configurable - N2 Equivalent
Standard MG-MR Bin
Configurations
SA40
SA41
SA42
SA43
SA44
SA45
SA46
SA47
SA48
SA50
The MultiFlo Configurator interfaces to the GF40/GF80 Series device
through RS485. There are various ways to connect the device regardless of
device configuration. Devices may be connected through the diagnostic port
using cables in one of the two Basic MultiFlo Configurator Kits.
778Z010ZZZ Basic MultiFlo Configurator Kit
A331710003 Cable Assembly 2.5mm
A332300001 Converter 232/485
Installation and Operation Manual
X-TMF-GF40-GF80-Series-MFC-eng
Part Number: 541B161AAG
Flow range
Gas Flow Range
Code
(N2 Equivalent)
010C
3-10 s ccm
030C
11-30 s ccm
092C
31-92 s ccm
280C
93-280 s ccm
860C
281-860 s ccm
2.6L
861-2600 s ccm
7.2L
2601-7200 s ccm
015L
7201-15000 s ccm
030L
15001-30000 s ccm
055L
30001-50000 s ccm
August, 2012