English
WARNING
Be certain to place the posterior section in a dry place until prepared to engage it with the anterior section.
n
Make certain all cables have remained in their respective connectors or modular jacks before attempting to assemble the EX3 Stepper.
n
ANTERIOR SECTION ONLY
CLEANING - IMMERSION AND WIPING ARE BOTH VALIDATED FOR CLEANING. CHOOSE EITHER METHOD.
IMMERSION
1. Prepare an enzymatic detergent solution, such as Enzol
2. Fully immerse anterior section of EX3 Stepper in prepared enzymatic detergent solution. Soak for 3 minutes.
3. After soak time, brush entire surface for 1 minute. Ensure all crevices and grooves are brushed.
4. Dispose of used enzymatic detergent solution and prepare a new enzymatic detergent solution, such as Enzol
manufacturer's recommendation.
5. Allow EX3 Stepper to soak in new enzymatic detergent for 3 minutes.
6. Remove EX3 Stepper from enzymatic detergent solution and rinse under running tap water for a maximum of 1 minute, but not less than 50
seconds.
7. Dry EX3 Stepper with a soft, clean, lint-free cloth.
NOTE:
If there is residual water inside the stepper, compressed air may be used at a maximum of 150 psi to remove remaining water.
n
Prolonged exposure of anodized aluminum and nylon to Enzol
n
WIPING
1. Using a clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal wipe,
wipe surfaces to remove gross soil.
Ensure all crevices and grooves have come into contact with wipe.
n
2. Using a new, clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal
wipe, wipe surfaces until visibly wet.
DISINFECTION
1. Unfold a clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth
and thoroughly wet surface.
2. Allow treated surface to remain wet for no more than 2 minutes. Use additional wipe or wipes, if necessary, to ensure continuous 2 minutes wet
contact time.
3. Once 2 minute wet contact time has been reached, allow surfaces to air dry.
POSTERIOR SECTION NON-ELECTRICAL COMPONENTS ONLY
Top Cover
n
Bottom Cover
n
Hold Screws
n
Longitudinal Movement Knob
n
WARNING
Do not soak or immerse the posterior section of EX3 Stepper containing electrical components.
n
CLEANING AND DISINFECTION
1. Using a clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal wipe,
wipe surfaces to remove gross soil.
Ensure all crevices and grooves have come into contact with wipe.
n
2. Using a new, clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal
wipe, wipe surfaces until visibly wet.
3. Allow treated surface to remain wet for no more than 2 minutes. Use additional wipe or wipes, if necessary, to ensure continuous 2 minutes wet
contact time.
4. Once 2 minute wet contact time has been reached, allow surfaces to air dry.
ASSEMBLING EX3 STEPPER AFTER REPROCESSING (REQUIRED IF IMMERSION CLEANING METHOD WAS USED)
1. Connect posterior section at 20 degree angle to anterior and rotate into place. Ensure location lock finger has fully engaged.
WARNING
Ensure interface screws are fully tightened to prevent gear slippage and corresponding errors in positioning when advancing the stepper.
n
®
Enzymatic Detergent, at manufacturer's recommendation.
®
Enzymatic Detergent may cause slight discoloration of surfaces.
6
EX3™ Stepper
®
Enzymatic Detergent, at
®
Germicidal wipe,