Set Instrument Parameters - Medtronic ACT Plus Manual Del Usuario

Ocultar thumbs Ver también para ACT Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
From the Main Menu:
1. Select [Transmit Test Results]. The following variable function keys will
appear: [Transmit All Patient Tests], [Transmit Unsent Patient Tests],
[Transmit By Patient ID], [Transmit All QC Tests], [Transmit Unsent QC
Tests], and [Exit to Main Menu].
2. Insert a 3.5-inch floppy disk, IF the floppy disk is selected.
3. To transmit all patient test results in the memory, select [Transmit All Patient
Tests].
4. To transmit the patient test results that have not previously been sent to the
selected output location, select [Transmit Unsent Patient Tests].
5. To transmit all results for a specific patient, select [Transmit By Patient ID].
Then enter the Patient ID and press [Enter].
6. To transmit all QC test results (liquid and electronic), select [Transmit All QC
Tests]. All QC tests in the memory will be sent to the selected output location.
7. To transmit all QC test results that have not been previously sent to the
selected output location, select [Transmit Unsent QC Tests].
8. To return to the Main Menu, press [Main Menu].

Set Instrument Parameters

Parameters for the ACT Plus™ instrument are set at the time of installation. This
must be completed prior to any other data entry.
1. Select [Instrument Parameters]. The following variable function keys
appear: [Date - yyyy-mm-dd], [Time – hh:mm], [Audio Tone], [Language],
[Go to Page 2 of 2], and [Exit to Main Menu].
2. To set the current date, select [Date - yyyy-mm-dd]. The following ranges are
acceptable for the year, month, and day entries: yyyy = year (accepted range,
1998 to 2097), mm = month (accepted range, 01 to 12), and dd = days
(accepted range, 01 to 31).
3. To confirm the date entry, press [Enter].
Note: The current date must be entered before cartridge and/or control lot
numbers and expiration dates are entered.
4. To set the current time, select [Time – hh:mm]. The current time is based on
a 24-hour clock where hh = hours (accepted range, 00 to 23) and
mm = minutes (accepted range, 00 to 59).
5. To confirm the time entry, press [Enter].
6. To set the audio tone to ON or OFF, select [Audio Tone].
7. To confirm the selection, press [Enter].
8. To select a language, select [Language]. The options are English, French
(Français), Italian (Italiano), German (Deutsch), or Spanish (Español).
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
Medtronic Confidential
CS010
English 23

Publicidad

Tabla de contenido
loading

Tabla de contenido