Set the QC Interval for Liquid and Electronic Controls
From the QC Manager Setup screen:
1. Select [QC Interval Setup]. The following variable function keys appear:
[Liquid Interval], [ACTtrac
To set the liquid QC interval:
1. Select [Liquid Interval].
2. To view the choices, [None], [8 hours] or [7 days], continue to press [Liquid
Interval].
3. To confirm the choice, press [Enter].
®
To set the ACTtrac
electronic QC interval:
®
1. Select [ACTtrac
2. To view the choices, [None] or [8 hours], continue to press [ACTtrac
Interval].
3. To confirm the choice, press [Enter].
Enable or Disable Test Cartridge Types
From page 2 of the QC Manager Setup screens:
1. Select [Cartridge Enable Setup].
2. To select the desired cartridge, use the arrows on either side of the display of
cartridges to navigate up or down.
3. Select the desired cartridge.
4. To view the choices, [ON] or [OFF], continue to press the desired cartridge.
5. To confirm the choice, press [Enter].
Download All of the Instrument Settings to a Floppy Disk
From page 2 of the QC Manager Setup screens:
1. Insert a formatted disk into the floppy drive.
2. Select [Download Settings].
Set the Permanent Record or the Mode by Which the
ACT Plus™ Instrument Maintains a Record of
Downloaded Data
From page 2 of the QC Manager Setup screens:
1. Select [Perm Record].
2. To view the choices, [None], [Floppy], or [Electronic], continue to press
[Perm Record].
32 English
86702-001
Rev 1.0
®
Interval] and [Exit to QC Manager Setup].