Location
[CVOR]
User ID Setup
Clear ID Interval
[20 mins]
The [Location] variable function key is used to select the location of the ACT
Plus™ instrument. Location options include the following: CVOR (cardiovascular
operating room), Cath Lab (Cardiac Catheterization Lab), ECMO (Extracorporeal
Membrane Oxygenation), Dialysis, ICU/CCU (Intensive Care/Coronary Care
Units), Lab, and Other.
The [User ID Setup] variable function key displays the screens for the following
functions: adding and deleting User IDs, turning ON or OFF User Lockout,
changing the QC Manager Password, downloading User IDs to a floppy disk, and
uploading User IDs from a floppy disk to the ACT Plus™ user database.
The [Clear ID Interval] variable function key is used to select the time interval
(none, 5 minutes, 10 minutes, or 20 minutes) that will elapse before the Patient and
User IDs are cleared; they will need to be reentered prior to testing.
The [QC Lockout] variable function key is used to set QC Lockout to one of the
following options: OFF, Warning, or ON.
The [QC Interval Setup] variable function key is used to set the required interval
for performing liquid and electronic QC tests. The options for liquid are the
following: None, 8 hours, or 7 days. The options for Electronic are 0 or 8 hours.
The [Cartridge Enable Setup] variable function key is used to enable or disable
the various cartridge types that may be run on the ACT Plus™ instrument.
The [Download Settings] variable function key is used to download all instrument
parameters and settings to a floppy disk.
The [Perm Record] variable function key designates the official mode of tracking
the results that have been downloaded. The options are None, Floppy, or
Electronic.
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
--->> QC Manager <<---
QC MANAGER SETUP
(Page 1 of 2)
Figure 6. QC Manager Setup Screen
Medtronic Confidential
CS010
QC Lockout
[Off]
QC Interval
Setup
Go to
Page 2 of 2
English 29