Message Priorities - Medtronic ACT Plus Manual Del Usuario

Ocultar thumbs Ver también para ACT Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
Information/Status: Screen message with no audio tone.
Alerts: Screen message with a three-beep audio tone (audio is active when
the [Audio Tone] parameter is set to ON).
System Alarms: Screen message with a single, long audio tone (audio is
always active). System Alarms are presented when the ACT Plus™
instrument's self-testing mechanism detects a condition that may affect the
ability of the instrument to function properly. These messages are displayed,
if possible, and may be accompanied by a 3-digit Error Code, which service
personnel use to identify the system error.

Message Priorities

Messages are displayed in a priority structure. If multiple error or precautionary
conditions occur simultaneously, only the highest-priority message appears on the
display.
.
System Messages
Both lot numbers are filled.
OR
Both cartridge lots full.
Bar code is not a cartridge type.
Bar code is not a control type.
Bar code is not valid.
Expired cartridge lot.
OR
Expired control lot.
Cartridge lot already exists.
OR
Control lot already exists.
Cartridge is not enabled.
Cartridge for control is not enabled.
Expiration date edited. QC due.
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
Medtronic Confidential
Cause/Resolution
The user is trying to add a lot number when
two already exist for the selected cartridge.
Remove one of the lot numbers.
The user is trying to scan a control lot on the
cartridge setup screen.
The user is trying to scan a cartridge lot on
the control setup screen.
The bar code information was not valid as a
cartridge or control. Rescan the bar code.
Reprogram the bar code scanner.
The user is attempting to run a test with a
cartridge/control that is beyond the
expiration date. Make sure that the expiration
date on the packaging matches the date in
the cartridge lot entry screen.
The user is trying to scan a lot number when
that lot number already exists for the
selected cartridge/control.
The cartridge selected is not enabled in the
QC Manager Setup screen. Contact your QC
Manager to enable the selected cartridge.
The user is trying to scan in a lot number for
a control when the cartridge that uses that
control is not enabled in the QC Manager
Setup. Contact your QC Manager to enable
the selected cartridge.
The expiration date for the cartridge lot has
been changed; control tests must be
repeated for this lot. Repeat the controls for
this lot number.
CS010
English 45

Publicidad

Tabla de contenido
loading

Tabla de contenido