Indications For Use - Medtronic MYOtherm XP Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36
MYOtherm XP™
Cardioplegia Delivery System with Cortiva™ BioActive Surface
1. Description
The MYOtherm XP™ cardioplegia delivery system with Cortiva™ bioactive surface is designed to mix arterial
blood from an oxygenator with asanguineous cardioplegia solution in specific ratios, depending upon the system
chosen. The specifications provided by each standard system and their corresponding product designations are
presented below in Table 1. The patient delivery line is packaged separately to allow for ease of sterile transfer
to the operative field prior to connection to the delivery outlet of the cardioplegia system. A pressure monitoring
line, three-way stopcock, gauge protector, and a pressure relief line with pressure relief valve are preconnected.
Products coated with Cortiva bioactive surface include a "CB" prefix in the model number. Additional information
regarding Cortiva bioactive surface is available in Section 7.
MYOtherm XP system with Cor-
tiva bioactive surface model
CBXP41, CBXP41B
a
Note: CBXP41B (Bridge) series provides the ability to administer 100% blood as prescribed by physician.
1.1. Specifications
Recommended blood flow rate
Filter screen
Heat exchanger nominal volume
Heat exchanger blood contact surface area
Size
Inlet connector
Outlet connector
Pressure monitor/pressure relief ports
Waterports
Water phase pressure limit
Blood phase pressure limit (postprime)
Maximum water temperature
Maximum transit temperature
Store the product at room temperature.
1.2. Accessories available separately
Temperature probe TP
MYOtherm XP™ cardioplegia holder

2. Indications for use

The MYOtherm XP with Cortiva bioactive surface is a device intended for the mixing, warming/cooling and
delivery of oxygenated blood/cardioplegia solution in a predetermined ratio during routine cardiopulmonary
bypass (CPB) procedures up to 6 hours in duration. Blood/cardioplegia solution is delivered to the patient
through the cardioplegia delivery system and appropriate cannula by the operation of a single occlusive roller
pump.
3. Contraindications
Use the device only as indicated.
4. Warnings
Read all warnings, precautions, and instructions for use carefully prior to use. Failure to read and follow all
instructions, or failure to observe all stated warnings, could cause serious injury or death to the patient.
Only persons thoroughly trained in CPB procedures should use this device. Operation of each device
requires constant supervision by qualified personnel for patient safety. Perfusion must be carefully and
constantly monitored.
Each device is intended for single use only. Do not reuse or resterilize. Sterilized using ethylene oxide.
The fluid pathway is sterile and nonpyrogenic. Inspect each package and device prior to use. Do not use the
device if the package is opened or damaged or the protective caps are not in place.
Attach tubing in such a manner as to prevent kinks or restrictions that may alter blood or water flow.
8
Instructions for Use
Delivery ratio
4:1
English
Table 1.
Tubing inner diameter dimensions Pump type
a
Blood line
Cardioplegia line
0.64 cm (1/4 in) 0.32 cm (1/8 in)
≤500 mL/min
150 μm
47 mL
0.06 m
2
(95 in
2
)
Adult
0.64 cm (1/4 in)
0.48 cm (3/16 in)
Standard female luer lock
Quick coupling connectors [1.3 cm (1/2 in)]
275.8 kPa (40 psi)
750 mmHg
42°C (107°F)
40°C (104°F)
Standard roller
pump

Publicidad

Tabla de contenido
loading