3.
Preparing for Spare Part Replacement
3.1. Tools Needed
•
Name plate removal tool
•
Plectrum tool
•
Screwdriver(Phillips #0) – Not supplied
•
Compatible (and functioning) Visualization Device – Not supplied
The name plate removal tool and the plectrum tool are provided with the spare parts kits (see section 1.5 Spare Parts).
3.2. Work Environment
CAUTION
Use ESD protection when disassembling, replacing any spare part, or reassembling the displaying unit.
It is recommended to work on a soft surface to avoid scratches on the screen and surface of the unit.
3.3. How to Disassemble the Displaying Unit
WARNING
To avoid risk of electric shock, always disconnect the mains from the displaying unit before performing spare
part replacement.
CAUTION
The displaying unit must be turned OFF before disassembly.
Follow these steps to disassemble the displaying unit:
a)
Clean the displaying unit before disassembly
WARNING
To avoid contamination, always clean and disinfect the displaying unit before and after the replacement and wear
gloves during handling. Follow the instructions provided in section 2 How To Clean and Disinfect the Displaying Unit.
b)
Remove all cables attached to the displaying unit
E.g. power supply, visualization devices and external equipment
c)
Remove the name plates (with the Ambu logo) on the top and bottom of the displaying unit
Refer to Figure 4: Insert the name plate removal tool in the hole at the end of the name plate. Press the name plate
removal tool down to filp the end of the name plate up. Use the tool as a lever. Gently lift the name plates upwards.
10
1
1
Figure 3
1
Name plate removal tool,
2
2
2
Plectrum tool
3