1.

Important information

This is a Spare Part Replacement Manual for Ambu® aView™ 2 Advance. It is applicable to aView 2 Advance with item
number 405011000 ONLY.
The following terms are used in this Spare Part Replacement Manual:
"Displaying unit" refers to Ambu aView 2 Advance.
"Visualization device" refers to the various Ambu scopes that can be connected and used with the displaying unit.
This manual may be updated without further notice. Contact your local Ambu representative or download the most
recent version from www.ambu.com.
Please observe that the warranty will be void, if the displaying unit has been disassembled within the warranty period
without the written consent from Ambu.
1.1. Requirements and Qualifications for Personnel Performing Spare
Part Replacement
Do not start replacing any spare part of the displaying unit, unless you have read through this document.
Do not start replacing any spare part of the displaying unit, unless you are qualified in the following areas:
Knowledge, experience and acquaintance with electronic repair and safety testing according to IEC 60601-1
(Medical electrical equipment, Part 1: General requirements for basic safety and essential performance) and
IEC 62353 (Medical electrical equipment – Recurrent test and test after repair of medical equipment).
Knowledge of and experience with local regulations.
Having read the Instructions for Use (IFU) for the displaying unit (download current version from www.ambu.com).
Knowledge of the environment where the displaying unit is installed and used.

1.2. Warnings and Cautions

WARNINGS
To avoid risk of electric shock, always disconnect the mains from the displaying unit before performing spare
pare replacement.
To avoid contamination, always clean and disinfect the displaying unit before and after the replacement and wear
gloves during handling.
To avoid compromising electrical safety, make sure not to leave any tools or other loose parts inside the displaying
unit when it is reassembled.
To avoid compromising electrical safety, always perform tests after the displaying unit has been reassembled.
If any part of the tests fails after spare part replacement, do not use the displaying unit.
Only use spare parts supplied by Ambu. Do not modify the spare parts.
For battery spare part kit: Risk of fire and burns. Do not open, crush, heat above 45 °C or incinerate the battery.
CAUTIONS
Use ESD protection when disassembling, replacing any spare part, or reassembling the displaying unit.
The displaying unit must be turned OFF before disassembly.
Before starting the replacement of any spare part, make sure that the battery is disconnected.

1.3. Data security

To avoid possible loss of data, it is recommended to export relevant files before performing any spare part replacement.
Refer to the online Reference Manual at www.ambu.com.