How To Clean And Disinfect The Displaying Unit - Ambu aView 2 Advance Manual Del Usuario

Ocultar thumbs Ver también para aView 2 Advance:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 177
2.

How To Clean and Disinfect the Displaying Unit

The displaying unit shall be cleaned and disinfected before and after performing replacement according to good medical
practice, described in the instructions below.
The displaying unit can be cleaned and disinfected using disposable germicidal hospital disinfection wipes designed for
hard, non-porous surfaces. Compatible disinfection wipes are based on Isopropyl alcohol 70-80%, Isopropyl alcohol/
Ammonium Chloride or Isopropyl alcohol/Ethanol.
a)
Turn OFF the displaying unit and make sure that it is unplugged from the power outlet.
b)
Follow the cleaning instruction from the germicidal disinfection wipe manufacturer to thoroughly clean the
displaying unit.
c)
Allow the displaying unit to dry after the cleaning.
d)
Follow the disinfection instruction from the germicidal disinfection wipe manufacturer to thoroughly disinfect the
displaying unit with a new wipe.
e)
Allow the displaying unit to dry after the disinfection.
f )
If the cleaning and disinfection chemicals used leave a residual layer on the display after they dry out, use an alcohol-
based wipe to remove the residue.
g)
If applicable, clean and disinfect the mounting bracket following the same procedure as for the displaying unit.
NOTES:
The wipes shall be moist, but not dripping to ensure no damage to internal electronics of the displaying unit.
If using wipes containing hypochlorite, ensure that all residue is completely removed. Wipes containing
hypochlorite may affect the screen's antireflective coatings over time. You should limit the use of wipes containing
hypochlorite to required cases only.