4.
How to Replace the Spare Parts
This section covers the replacement of different parts. You may jump directly to the section covering the specific part to
be replaced:
•
Battery (section 4.1)
•
Visualization device interface (VDI) board (section 4.2)
•
Fan (section 4.3)
•
USB cover (section 4.4)
WARNING
Only use spare parts supplied by Ambu. Do not modify the spare parts.
4.1. Battery Replacement
Ambu® aView™ 2 Advance - Battery Kit
Battery
Replacement part kit:
1
Name plates (x2)
2
Screw cover (x2)
3
Visualization device connector cover (x2)
4
Plectrum tool (x1)
5
USB Cover (x1)
6
Screws for casing assembly (x10)
7
Screws for VESA the mounting bracket (x4)
8
Name plate removal tool (x1)
WARNING
Risk of fire and burns. Do not open, crush, heat above 45 °C or incinerate the battery.
Replacing Battery
Refer to Figure 12 to identify the location of the battery.
a)
Pull the battery off the displaying unit main board
The battery is fastened with double-sided adhesive tape.
405012100
1
1
4
3
3
2
2
7
7
6
6
4
5
5
8
8
15