5 What if ...?
Indicator
Causes
The device is overheated.
Red "x" lights up
Electronic component of the
handpiece is defective.
Battery empty
Red "x" and
battery symbol
light up
Battery contacts dirty
In the 3sCure program, the
3s flashes
curing time is set to 3 seconds
and cannot be changed. The
3sCure program can be used no
more than twice in direct
succession.
The charging base
– Power pack not connected or
is not illuminated
defective
– Battery fully charged
during charging
16
Error rectification
Allow the device to cool down and try again after a
certain time. If the error persists, please contact your
dealer or your local Service Centre.
Remove and reinsert the battery. If the error persits,
please contact your dealer or your local Service Centre.
Place the device in the charging base and charge it.
Remove battery and clean the battery contacts.
After two consecutive curing cycles, an extended
waiting time of 30 seconds is necessary for safety
reasons before the next curing cycle can be started.
If for compelling reasons an additional curing cycle is
required before the prescribed interval of 30 seconds
has elapsed, the 3sCure program can be activated
again by selecting it anew with the program selection
button.
Check if the power pack is correctly positioned in the
charging base or if the power pack is connected to the
power supply by means of the power cord.