Edge trigger
The Edge trigger identifies a trigger by looking for a specified edge (slope) and
voltage level on a waveform.
Source
Press
Coupling
Press
– DC coupling – allows DC and AC signals into the trigger path.
– AC coupling – removes any DC offset voltage from the trigger waveform.
– LF (low frequency) reject coupling – removes any unwanted low-frequency
– HF (high frequency) reject coupling – removes high-frequency components
Slope
Press
Alternate ( ) edges, or Either ( ) edge.
All modes operate up to the oscilloscope bandwidth except the Either edge mode,
which has a limitation. The Either edge mode will trigger on constant wave signals
up to 100 MHz, but can trigger on isolated pulses down to 1/(2 × oscilloscope
bandwidth).
Level
Press
Keysight U1610/20A User's Guide
repeatedly to select the trigger source.
repeatedly to select either:
components from the trigger waveform.
from the trigger waveform.
repeatedly to select the Rising ( ) edge, Falling ( ) edge,
and use the
Using the Oscilloscope
or
key to set the trigger level.