Descargar Imprimir esta página

Elcometer NDT MTG6 Guia Del Usuario página 18

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 121
8 CALIBRATING YOUR GAUGE (continued)
8.4 USING VELOCITY CALIBRATION
To calibrate the gauge using this method, the user must know the
sound-velocity of the test material. This calibration method is useful if
uncoated sample test pieces of known thicknesses are not available.
1
Press the On/Off button to switch the gauge on.
2
Press Menu/Calibration/Cal Method and select 'Velocity'.
If 'Velocity' is already selected - the calibration method currently selected
is indicated by the icon to the right of the display - simply press
Menu/Calibration/Calibrate.
3
Enter the known sound-velocity using the
0 to 9 and the  softkey to move to the next digit, followed by
'Set' to use the value entered.
Pressing 'Escape' at any time will exit the calibration procedure without
calibrating the gauge.
The sound-velocity entered will be displayed to the right of the display,
below the calibration method icon.
8.5 USING THICKNESS SET CALIBRATION
Only available when in 'Velocity Mode' - see Section 4.6 'Selecting the
Measurement Mode' on page en-8 - to calibrate the gauge using this
method, the thickness of the test material must be known.
1
Press the On/Off button to switch the gauge on.
2
Press Menu/Calibration/Cal Method and select 'Thickness Set'.
If 'Thickness Set' is already selected - the calibration method currently
selected is indicated by the icon to the right of the display - simply press
Menu/Calibration/Calibrate.
3
The user will be prompted to perform a 'Zero Set' to set the zero
point of the transducer, which is recommended before calibrating
the gauge - see Section 6 'Setting the Zero Point' on page en-12.
4
Enter the known material thickness using the
select 0 to 9 and the  softkey to move to the next digit, followed
by 'Set' to use the value entered.
Pressing 'Escape' at any time will exit the calibration procedure without
calibrating the gauge.
The material thickness entered will be displayed to the right of the display,
below the calibration method icon.
en-17
R
softkeys to select
softkeys to
www.elcometer.com

Publicidad

loading

Este manual también es adecuado para:

Ndt mtg8