8 CALIBRATING YOUR GAUGE (continued)
To enable and disable calibration check:
1
Press Menu/Calibration.
2
Use the
'Select'.
3
To disable, press 'Select' again to un-check the 'Calibration
Check' radio button.
8.9 LOCKING THE CALIBRATION
Using the 'PIN Lock' feature, the calibration settings can be 'locked',
preventing the user from making any changes to the calibration
without first disabling PIN lock.
Users can still test the calibration via Menu/Calibration/Test Calibration
when 'PIN Lock' is enabled, but are unable to validate or re-calibrate
the gauge.
For more information on 'PIN Lock', see Section 9 'PIN Lock' on page
en-20.
8.10 CALIBRATION MEMORIES - MTG8
Up to three calibrations can be saved in the gauge memory. Once
saved, the user can select the calibration memory - without the need
to re-calibrate the gauge.
To save a calibration into memory:
1
Press Menu/Calibration/Cal Memory n, where n = 1, 2 or 3.
2
Use the
'Select'.
3
Use the
and follow the on-screen instructions to calibrate the gauge.
4
The calibration will be stored in the gauge memory as Cal
Memory n, where n = 1, 2 or 3.
To rename a calibration memory, press Menu/Calibration/
Cal Memory n/Rename Cal Memory n.
To view the calibration memory data, press Menu/Calibration/
Cal Memory n/View Calibration Data.
en-19
R
softkeys to highlight 'Calibration Check' and press
softkeys to highlight 'Cal Method' then press
softkeys to highlight the required calibration method
www.elcometer.com