D. Adjust the water bottle hold-down bracket to fit the water bottles used by your facility. This
is done by loosening the nut on the back and sliding the bracket up or down so it contacts the
water bottle and then re-tightening the nut.
6.
Plug the water warmer cable into the receptacle marked "water warmer" on the back of the unit.
Air hose attachment
Tools required: (not included)
9/16" open-end wrench
¾" open-end wrench
Accessories required:
"D" or "E" size CO
High pressure hose
Yoke adapter
1.
Assemble the high pressure hose and the yoke by tightening the hose with the 3/4" wrench while
holding the yoke with the 9/16" open-end wrench.
2.
Unscrew the T-handle on the yoke to allow it to slip over the top post of the CO
orientation is so that the two pins on the yoke line up with the two alignment holes on the tank.
(CO
tank is not included with system, purchase separately)
2
3.
After aligning the two pins, tighten the T-handle firmly.
4.
Tighten the other end of the high pressure hose and yoke assembly to the back of the
ENDO STRATUS® CO
tighten the hose nut. DO NOT USE TEFLON TAPE OR THREAD SEALANT ON THE FITTINGS.
Pre-Procedure setup
1.
Open the valve on the CO
2.
Turn the power switch to "On"
3.
Test the setup by turning the flow control knob position clockwise. Verify that there is CO
out the front output of the unit. If there is no CO
CO
tank is full and check the connection outlined in air hose attachment above.
2
4.
Turn off the flow control knob.
5.
Connect the tubing to the front output of the device. Use only ENDO SMARTCAP
ENDOGATOR
®
6.
When using ENDO SMARTCAP™ CO
ENDO SMARTCAP tubing or ENDOGATOR hybrid tube set. Attach ENDO SMARTCAP tubing or
ENDOGATOR hybrid tube set to the GI endoscope according to the manufacturer's instructions.
IFU-52 / REV H
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 12
tank filled with medical grade CO
2
insufflator. Hold the brass fitting with wrench and use a second wrench to
2
tank approximately 1 turn.
2
hybrid tube set manufactured by MEDIVATORS.
source tubing, attach ENDO SMARTCAP source tubing to
2
(not included)
2
observed out the front of the unit verify that the
2
12
tank. The correct
2
flowing
2
tubing or
®
12/11/2015 9:31:57 AM