Certifications:
Classification:
IP rating (Ingress Protection):
WARNING:
Grounding reliability can only be achieved when connected to a receptacle marked "Hospital Grade".
Mechanical Specifications
Physical Dimensions:
Height
Width
Depth
Weight
CO
Specifications
2
Free Flow rates (maximum unrestricted output to GI endoscope system):
Front panel setting
Low:
Medium:
High:
GI Endoscopic Flow rates (approximate flow rate at GI endoscope output):
Front panel setting
Low:
Medium:
High:
*These values are approximate and are based on testing with respective models of GI endoscopes. User results
may vary depending on GI endoscope used, channel diameter, and channel length.
Output Connection:
Input Connection:
CO 2 Heater:
CO 2 Output temperature:
Operating Pressure:
Input Pressure
*The low pressure warning light will illuminate at 25 PSI or less, although the unit may still flow. It is
recommended to replenish the CO
Pressure regulator set point
Internal safety relief valve
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 7
IEC-60601-1, IEC-60601-1-2, IEC-60601-2-18
Class 1 Type B
IP24
4 ¾" inches
7 ¾" inches
13 ¾" inches
10.5 pounds
1.4 L/min
2.4 L/min
3.5 L/min
0.6-0.9 L/min*
1.2-1.4 L/min*
1.5-1.8 L/min*
Male luer lock
¼" male flare fitting.
20W, internal, flow through, with dual redundant temperature sensors.
37°C (+3°C)
1900 PSI Maximum
25 PSI Minimum*
supply when the light illuminates.
2
8 PSI Maximum
≤12 PSI
121 mm
197 mm
349 mm
4.8 kg
7
IFU-52/ REV H