Table 4 – Guidance And MANUFACTURER'S Declaration
For ME EQUIPMENT And ME SYSTEMS That Are Not LIFE-SUP-
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The EGA-501/EGA-501E is intended for use in the electromagnetic environment speci-
fied below. The customer or user of the EGA-501/EGA-501E should ensure that it is
used in such an environment.
Immunity Test
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC 61000-4-3
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 17
Electromagnetic IMMUNITY
PORTING
Compliance
EN/IEC 60601
Level
Test Level
3 Vrms
(3)Vrms)
150 kHz to 80 MHz
3 V/m
(3)V/m)
80 MHz to 2.5 GHz
Electromagnetic Environment – Guidance
Portable and mobile communications equipment
should be separated from the EGA-501/EGA-501E
by no less
than the distances calculated/listed below:
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
Where P is the max power in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as
determined by an electromagnetic site survey, should
be less than the compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment
containing a transmitter.
17
IFU-52/ REV H