Table 2 – Guidance And MANUFACTURER'S Declaration
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The EGA-501/EGA-501E is intended for use in the electromagnetic environment specified below. The
customer or user of the EGA-501/EGA-501E should ensure that it is used in such an environment.
Immunity Test
ESD
EN/IEC 61000-
4-2
EFT
EN/IEC 61000-
4-4
Surge
EN/IEC 61000-4-5
Voltage Dips/
Dropout
EN/IEC 61000-
4-11
Power Frequency
50/60Hz
Magnetic Field
EN/IEC 61000-
4-8
Note 1 The EUT shuts off and must be manually restarted by the operator during a 5 second loss of AC Mains
power.
IFU-52 / REV H
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 16
Electromagnetic IMMUNITY
For All ME EQUIPMENT And ME SYSTEMS
EN/IEC 60601
Compliance
Test Level
Level
±6kV Contact
±6kV Contact
±8kV Air
±8kV Air
±2kV Mains
±2kV Mains
±1kV I/Os
±1kV I/Os
±1kV Differential
±1kV Differential
±2kV Common
±2kV Common
>95% Dip for
100% Dip for
0.5 Cycle
0.5 Cycle
60% Dip for
60% Dip for
5 Cycles
5 Cycles
30% Dip for
30% Dip for
25 Cycles
25 Cycles
>95% Dip for
Note 1
5 Seconds
3A/m
3A/m
Electromagnetic Environment – Guidance
Floors should be wood, concrete or ceramic tile. If
floors are synthetic, the r/h should be at least 30%
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment. If the
user of the EGA-501/EGA-501E requires continued
operation during power mains interruptions, it is
recommended that the
EGA-501/EGA-501E be powered from an uninter-
ruptible power supply or battery.
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.