proxeo
AURA
The device packaging is sensitive to impact,
therefore caution must be exercised during its
transport and deposit.
Do not stack multiple boxes so as not to crush
the underlying packages.
All material sent by Mectron is checked at the
time of dispatch.
The device is dispatched with suitable protec-
tion and packaging.
4
INSTALLATION
The air polisher device, in order to operate
must be connected to a dental unit equipped
for water and air distribution.
NOTE: The device does not require any
electrical and/or electronic connections to be
made.
The air polisher device is supplied with a specific
adapter for the main couplings to the dental unit
turbine (see Chapter 4.1 on page 11
CAUTION: Use the air polisher device
only with the specific coupling for the relative
turbine.
Connecting the air polisher device to
a connector not intended for use in
combination with the dental unit may
damage the device.
The dental unit to which air polisher is connect-
ed must be a medical device that complies with
general and applicable product regulations for
the specific category of dental unit, and in gen-
eral, satisfy the following standards:
• ISO 7494 Dentistry – Stationary dental
units and dental patient chairs;
• ISO 9168 Dentistry – Hose connectors
for air driven dental handpieces;
• ISO 60601-1 Medical electrical equipment
– Part 1: General requirements for basic
safety and essential performance.
To ensure the correct operation of the device, it is necessary to check:
• The "spray" water supply coming from
the turbine;
• The air supply coming from the turbine;
• The presence of powder.
10
When the device is received, check the condi-
tion of the packaging and in case of damage,
deterioration, or signs of tampering, submit a
claim to the carrier.
Keep the packaging in case any items need to
be sent to an Authorised W&H Service Centre
and to store the appliance during long periods
of non-use.
CAUTION: Check that the dental unit to
which air polisher is connected satisfies the
specified requirements.
The device must be installed in a suitable
place allowing its comfortable use.
For the air polisher device to operate correct-
ly, the air and water pressure values must be
within the intended range of use, specifically:
• For water, from 1.0 to 3.0 bar (from
)
.
1000 hPa to 3000 hPa), max 40 °C;
• For air, from 2.2 to 3.0 bar (2200 hPa -
3000 hPa).
CAUTION: To ensure proper and long-
lasting performance, ensure the air flowing
from the dental unit is thoroughly dry and
filtered.
Moreover, it is essential to ensure that the air
pressure is between 2.2 and 3.0 bar inclusive,
measured at the point of the connection
between the device and the dental unit.
NEVER use excessive air pressure.
Increasing the air pressure does not
correspond to an increase in performance;
to the contrary, the device may become
damaged or fail to work properly.