CLEANING, DISINFECTION AND STERILISATION
6
CLEANING, DISINFECTION AND
STERILISATION
The following tables are for indicative purpos-
es only.
For the full procedures for cleaning and sterili-
sation of individual parts refer to the Chapters
referenced in the tables.
CAUTION: It is prohibited to use methods
not covered in the tables below.
Phase
Chapter
I
6.1
Preparation
II
6.2
Manual cleaning with detergent solution
III
6.3
Disinfection with isopropyl alcohol-based disinfectant
IV
6.4
Daily decontamination of the water circuit of the dental unit
NOTE: To clean the accessories, select a
procedure (manual or automatic) from
among those validated and proposed in this
manual.
Accessories (Polisher Nozzles, K9 Wrench, Cleaning Needle)
Phase
Chapter
6.5.1
Manual cleaning
V
6.5.2
Automatic cleaning
VI
6.6
Check cleaning
VII
6.7
Drying
VIII
6.8
Sterilisation
NOTE: Repeated reconditioning has a
minimal effect on the devices and their
accessories. The end of the service life of
the accessories is generally determined by
wear or damage resulting from use. Mectron
guarantees the integrity of its sterilisable
polisher nozzles for up to 250 reconditioning
cycles.
operations described below must be
carried out at the time of first use and at all
subsequent uses.
TIPs are single-use sterile.
The SUBGINGIVAL PERIO TIPs are supplied
sterile and for single use. They must be
used on one patient only during a single
application, after which they must be
removed and disposed of.
air polisher device
Polisher
Procedure
Nozzles
CAUTION: The cleaning and sterilisation
WARNING: The SUBGINGIVAL PERIO
Procedure
K9 Wrench
x
x
x
x
x
x
x
x
x
x
Cleaning
Needle
x
x
x
x