Polisher Nozzles
Free the channel of the polisher nozzle of
any powder residue using the supplied Ø
0.4 mm cleaning needle, from both sides.
Blow compressed air into the central hole
in the tip, from both ends;
CAUTION: Clean the channel of the
nozzle using only the Ø 0.4 mm cleaning
needle supplied with the appliance.
CAUTION: If the SUBGINGIVAL
PERIO TIP is on the Polisher nozzle,
remove and dispose of it.
Place the polisher tip horizontally in a
clean container and add enzymatic de-
tergent solution until the polisher tip is
covered completely;
Leave to soak for 10 minutes at 22 °C ±2
°C. This procedure reduces the amount
of blood, protein and mucous present on
the polisher nozzle;
After 10 minutes of immersion in the en-
zymatic solution, gently brush the surface
of the polisher nozzle using the brush
with soft nylon bristles for at least 20
seconds, and eliminate all visible residue.
Thoroughly clean the hard-to-reach are-
as such as the edges, recesses, and joints;