proxeo
AURA
Dry the air polisher device with a clean,
soft, and low-lint cloth;
CAUTION: Do not sterilise air polisher in
an autoclave. It may be damaged and cause
personal injury and/or property damage.
CAUTION: Do not immerse the
air polisher device in various liquids and/or
solutions as it may be damaged.
6.3
Disinfecting the air polisher Device
The following procedure must be carried out on the air polisher device.
CAUTION: DO NOT USE as disinfectants::
• Very alkaline products (pH > 9);
• Products containing sodium
hypochlorite;
• Products containing hydrogen peroxide;
• Products containing abrasive substances;
• Very acid products (PH < 4);
• Products containing aldehyde, amine
and/or phenols;
6.3.1
Required Materials
• Disinfectant made with isopropyl alcohol
(> 60% v/v) and quarternary ammonium
compounds
;
b)
• Clean, dry, and low-lint cloths
24
4
CAUTION: Do not immerse the
air polisher device in the ultrasonic tank.
CAUTION: Do not use running water to
clean the air polisher device.
• Acetone;
• Methylethylketone;
since they may discolour and/or damage the
plastic materials.
The manufacturer disclaims all liability for any
damage caused by the substances mentioned
above. In case of damage caused by those
substances, the Warranty will be void.
f) Disinfectant with tuberculocidal declaration (pH 6-9).
The disinfection procedure has been validated by an
independent body using the following disinfectants
for surfaces: Discide Ultra - Disinfecting Spray -
.
Manufacturer: Palmero Healthcare LLC.
c)
g) The disinfection procedure has been validated by
an independent body using the following product:
Discide Ultra Soaked Wipes - Manufacturer: Palmero
Healthcare LLC.