CLEANING, DISINFECTION AND STERILISATION
While soaking in the enzymatic solution,
gently brush all the surfaces until all visi-
ble soil has been removed.
Use a clean brush with soft nylon bristles
for the external surfaces, and a clean pipe
cleaner with soft nylon bristles for the in-
ternal cavities and grooves.
NOTE: Thoroughly brush all the
following parts of the K9 wrench for
about 20 seconds:
• Through-holes and internal
channels;
• Hard-to-clean areas such as
sharp edges and in particular, the
interstices between the cutting
cusps;
• External metal ring;
• Internal cavities, grooves and
fissures.
Remove the K9 wrench from the enzy-
matic detergent solution.
Thoroughly rinse and brush all the sur-
faces of the K9 wrench (see previous
point) under running water for at least 10
minutes;
Place the K9 wrench in the ultrasonic tank
immersed in the enzymatic detergent
solution at 24 °C ±2 °C and run a cycle
for at least 20 minutes;