Descargar Imprimir esta página

Keysight DAQ970A Guia Del Usuario página 112

Publicidad

Idiomas disponibles

Idiomas disponibles

3     Features and Functions
5. Press Gage Settings, and use the knob or the front panel arrow keys to specify various strain gage settings.
Softkey
Gage Factor
Excitation
Fix/ Ext
Excitation
Voltage
6. Press Integrate Settings softkey to select whether the measurement integration time will be specified in terms
of Number of Power Line Cycles (NPLC) or directly in seconds (Time). Only 1, 2, 10, 20, 100, and 200 PLC
provide normal mode (line frequency noise) rejection. Selecting 200 PLC provides the best noise rejection and
resolution, but the slowest measurements.
Softkey
Integrate
NPLC
Time
112
Description
Gage factor is the ratio of the fractional change in resistance to the fractional change in length (strain)
along the axis of the gage. The larger the value, the more sensitive strain gage. Gage factor is a
dimensionless quantity. The typical value is approximately 2.
Strain bridge conversions require the voltage of the external bridge excitation. For this voltage, you
can dedicate a multiplexer channel to measure the excitation voltage, or can specify a known fixed
voltage value. Fixed (
Fix ) - The fixed value specified by the Excitation Voltage softkey will be used for
the strain conversion. External (
used for subsequent strain bridge measurements that specify an external excitation voltage source.
Note that the external DCV reference channel must be a lower-numbered channel than the strain
channel.
Use the knob or the front panel arrow keys to specify the fixed excitation voltage applied to the bridge
by an external voltage source. This value will be used to convert strain bridge measurements on the
selected channel.
Description
Sets the integration time as 0.001, 0.002, 0.006, 0.02, 0.06, 0.2, 1, 2, 10, 20, 100, and 200 Power Line
Cycles (PLC). Only 1 PLC and longer times provide normal mode (line frequency noise) rejection.
Selecting 100 or 200 PLC provides the best noise rejection and resolution.
Directly sets the integration time in seconds.
Ext ) - DCV measurements on the enabled reference channel will be
Keysight DAQ970A/DAQ973A User's Guide

Publicidad

loading

Este manual también es adecuado para:

Daq973a