Descargar Imprimir esta página

Keysight DAQ970A Guia Del Usuario página 64

Publicidad

Idiomas disponibles

Idiomas disponibles

3     Features and Functions
[Scan/Start] Key
Press the [Scan/Start] key on the front panel to initiate any scanning or digitizing operations, or to perform a single
sweep/record when operating in manual trigger mode. See the
[Scan/Start] key is used when the manual trigger mode is selected.
The [Scan/Start] key's backlight is turned on while scanning/digitizing is in progress. To stop the scan or digitizer
operation, press and hold the [Scan/Start] key for more than 2 seconds. In Scan data acquisition mode, "Scan
stopped" is shown when the operation is terminated. Likewise, in DMM Digitize and Digitizer data acquisition mode,
"DMM Digitize stopped" and "Digitizer stopped" will be shown respectively when the operation is terminated.
When the Auto Save feature in the [Save Recall] menu is turned on for DMM Digitize/Digitzer data acquisition
mode, the digitizing operation will not begin unless the instrument detects a USB drive connected to the USB
port at the front panel.
64
[Interval] Menu
for more information on how the
Keysight DAQ970A/DAQ973A User's Guide

Publicidad

loading

Este manual también es adecuado para:

Daq973a