Scheduled Autocalibration
You can also schedule the instrument to perform autocalibration based on a scheduled time. To use this feature,
press Scheduled Autocal > Schedule > On. To specify the frequency, press Interval > Daily/ Weekly/ Monthly.
Softkey
Interval
Daily
Weekly
Monthly
Calibrate
Calibrate accesses the instrument calibration procedure. See Keysight DAQ970A/DAQ973A Service Guide for
further details.
Keysight DAQ970A/DAQ973A User's Guide
Description
1. To schedule the instrument to perform daily autocalibration, press
2. To set the time for daily autocalibration, press the
knob or front panel arrow keys to set your desired time of the day. Once complete, press
1. To schedule the instrument to perform weekly autocalibration, press
2. To set the day and time for weekly autocalibration, press the
softkeys respectively and use the knob or front panel arrow keys to set your desired day of the week and
time of the day. Once complete, press
1. To schedule the instrument to perform weekly autocalibration, press
2. To set the day and time for monthly autocalibration, press the
respectively and use the knob or front panel arrow keys to set your desired day of the month and time of
the day. Once complete, press
Hour , or the Minute softkeys respectively and use the
Week Day, the Hour , or the Minute
Done .
Done .
3 Features and Functions
Interval > Daily
Done .
Interval > Weekly
Interval > Monthly
Day, the Hour , or the Minute softkeys
175