3     Features and Functions
2-wire and 4-wire Thermistor
This feature is only available when Scan is selected as the acquisition mode.
1. Connect a thermistor to the module's screw terminals.
2. Press [Channel] > Measure, and select TEMP from the selection menu.
3. Press the Units softkey to specify the temperature measurement units as °C (Celsius), °F (Fahrenheit), or K
(Kelvin). You can mix temperature units on different channels within the instrument and on the same module.
4. Press Sensor Settings > Sensor and select Thermistor. Next, toggle the 2-Wire/4-Wire softkey to specify the
type of transducer probe used as 2-wire or 4-wire Therm. After that press the Type softkey to select the type of
thermistor. Supported types are 2.2K, 5K (default), and 10K.
5. Toggle Use as ref to enable (On) or disable (Off) the currently selected channel to be used as the reference
channel for subsequent thermocouple measurements that specify an external reference source.