AC Current
The AC Current measurement option will only be available when Scan is selected as the acquisition mode.
AC Current measurement is supported only on DAQM901A's channels 21 and 22.
This section describes how to configure the selected channel for AC current measurements from the front panel.
1. Connect a current source to the module's screw terminals.
2. Press [Channel] > Measure, and select ACI from the selection menu.
3. Press Range and use the knob or the front panel arrow keys to specify the measurement range. Auto
(autorange) automatically selects the range for the measurement based on the input. Autoranging is
convenient, but it results in slower measurements than using a manual range. Autoranging goes up a range at
120% of the present range, and down a range below 10% of the present range.