7
Ventilation openings
The ventilation openings in the housing provide adequate air circulation for dissipating heat
generated by the electronics. It is important to ensure that these openings are never covered.
8
Connection for Circuplodes
The electrode is connected to this connection via the HF cable. The length and impedance of this
cable are of critical importance. For this reason, only use the cable supplied by Enraf-Nonius.
Connection of any cable other than the one supplied for the purpose by Enraf-Nonius can
adversely affect the safety of the patient and the correct functioning of the equipment and
is therefore not permitted.
9
Attachment points for the electrode arms (4x)
10
Mains input with fuses
The mains cable supplied is connected to this socket, which is fitted with two mains fuses. For
replacement of defective fuses, see chapter "Fault conditions".
11
Connection for potential equalization cable
This connection is used in areas where local regulations require connection of a potential
equalization cable. A special cable can be supplied for this purpose. Please refer to our
catalogue.
12
Type number plate
This plate provides all information for identifying the equipment, such as type, version and serial
numbers (for service, guarantee etc.) together with connection data such as mains voltage and
current consumption.
EN
5