Power - Enraf Nonius Curapuls 670 Manual De Uso

Ocultar thumbs Ver también para Curapuls 670:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

7 Power

Pulse power (Pp)
The pulse power is the peak power of
the pulse, i.e. the height of the pulse as
shown in the diagram at the right.
Mean power (P)
The mean power is determined by the
pulse power, the pulse duration and the pulse frequency.
The mean power resulting from the parameters set can be displayed by switching from the pulse power
selector to the mean power selector.
The figure below indicates the mean power at different pulse frequency and pulse width at a puls power
of 200 Watt. The grey columns indicate the values with use of Circuplode-E. The grey columns are not
applicable for the Circuplodes 90 and 140 mm. These columns indicates additional values for the
Circuplode-E.
26
μs \ Hz
65
0,34
82
0,42
110
0,58
150
0,78
200
1,04
300
1,56
400
2,08
EN
6
35
46
62
0,46
0,60
0,80
0,58
0,76
1,02
0,78
1,02
1,36
1,06
1,38
1,86
3,40
1,84
2,48
2,10
2,76
3,72
2,80
3,68
5,12
82
110
150
1,06
1,44
1,96
1,34
1,80
2,46
1,80
2,42
3,30
2,46
3,30
4,50
3,28
4,40
6,00
4,92
6,60
9,00
6,56
8,80
12,00
200
300
400
2,60
3,90
5,20
3,28
4,92
5,56
4,40
6,60
8,80
6,00
9,00
12,00
8,00
12,00
16,00
12,00
18,00
24,00
16,00
24,00
32,00
600
800
7,80
10,40
9,84
13,10
13,20
17,60
18,00
24,00
24,00
32,00
36,00
48,00
48,00
64,00

Hide quick links: