Ensure that neither solid particles nor liquids can penetrate into the product.
CAUTION
Changing Axon-Bus gripping components when switched on
Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.
►
Power down the Axon-Bus prosthetic system by pressing the button in the charging recept
acle before changing any Axon-Bus components (e.g. Axon-Bus gripping component).
CAUTION
Unintentional unlocking of the Axon-Bus gripping component
Injury due to releasing the Axon-Bus gripping component from the forearm (e.g. while carrying
objects).
►
The two release buttons should only be activated in order to change the Axon-Bus gripping
component, deliberately and with due consideration of the respective situation.
CAUTION
Storing the 8E500 Michelangelo Hand in the closed position
Injury due to faulty control or malfunction because of damage to the sensors and mechanism.
►
Store the 8E500 Michelangelo Hand in the open state (neutral position) only.
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
►
Only clean the product with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der
maClean).
Use only the following products for cleaning/disinfecting the inner socket:
►
Cleaning: 453H10=1 Ottobock DermaClean
Disinfection: 453H16 Ottobock DermaDesinfect.
INFORMATION
Switching off the Axon-Bus prosthetic system during extended passive pauses (e.g. air or rail
travel, visit to a theatre or cinema, etc.) will make the battery last longer before requiring a
recharge. Only the entire Axon-Bus prosthetic system with all connected components can be
switched off. Individual Axon-Bus components cannot be switched off separately.
4.7 Information on the Power Supply/Battery Charging
CAUTION
Charging the product with soiled or damaged contacts
Injury due to unexpected behaviour of the product caused by insufficient charging.
►
Make sure that the contacts are always clean and free of grease.
►
Clean the electrical contacts of the charging plug and charging receptacle regularly using
cotton swabs and a mild soap solution.
Take care to avoid damaging the contacts with pointed or sharp objects.
►
NOTICE
Use of incorrect power supply unit/battery charger
Damage to product due to incorrect voltage, current or polarity.
19