Étiquettes et symboles
VISIBLE AND INVISIBLE LASER RADIATION
Avoid eye or skin exposure to direct or scattered radiation.
Class 4 laser product.
Maximum Power
Pulse duration
Wavelengths
Aiming beam (Class 3R)
Standard IEC EN 60825-1: 2007
LASER RADIATION
AVOID EYE OR SKIN EXPOSURE TO DIRECT OR
SCATTERED RADIATION
CLASS 4 LASER PRODUCT
CAUTION – CLASS 4 LASER RADIATION
WHEN OPEN AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
ATTENTION – HIGH VOLTAGE
THE CAPACITORS CAN BE CHARGED AT A
POTENTIALLY DANGEROUS ELECTRIC VOLTAGE
BEFORE REMOVING THE PROTECTIONS:
- SWITCH OFF POWER SUPPLY
- WAIT FOR 5 MINUTES
- CHECK FOR RESIDUAL VOLTAGE ON CAPACITORS
AVERTISSEMENT—Rayonnement
laser visible et invisible.
Arrêt du laser
Cook Medical © 2012
30W
150-800 µs
2100 nm
< 5 mW @ 532nm
Ouverture du laser
Pédale
dispositif d'activation du laser
Retourner à la Table des matières
RAYONNEMENT LASER VISIBLE ET INVISIBLE
Éviter l'exposition des yeux ou de la peau au rayonnement
direct ou diffus. Produit laser de classe 4.
Puissance maximale
Durée d'impulsion
Longueurs d'onde
Faisceau de visée (Classe 3 R)
Norme CEI EN 60825-1: 2007
RAYONNEMENT LASER
ÉVITER L'EXPOSITION DES YEUX OU DE LA PEAU AU
RAYONNEMENT DIRECT OU DIFFUS
PRODUIT LASER DE CLASSE 4
MISE EN GARDE – RAYONNEMENT LASER DE CLASSE 4
LORSQU'IL EST OUVERT, ÉVITER L'EXPOSITION DES
YEUX OU DE LA PEAU AU RAYONNEMENT DIRECT
OU DIFFUS
ATTENTION – HAUTE TENSION
LES CONDENSATEURS PEUVENT ÊTRE CHARGÉS
AVEC UNE TENSION ÉLECTRIQUE POTENTIELLEMENT
DANGEREUSE. AVANT LE RETRAIT DES PROTECTIONS :
- METTRE L'APPAREIL HORS TENSION
- PATIENTER 5 MINUTES
- CONTRÔLER L'ABSENCE DE TENSION RÉSIDUELLE
SUR LES CONDENSATEURS
Applicateur de fibre optique
ALLUME
ETEINT
Chapitre 1
Introduction
30 W
150 à 800 µs
2100 nm
< 5 mW à 532 nm
8-9