TROUBLESHOOTING
The DolorClast smart provides an error code system in case of malfunction or incorrect handling:
1 LED blinking = Handpiece error; 2 LEDs blinking = Air pressure error; 3 LEDs blinking = Software error
Symptom
The display does not light up when the
unit is switched on
The display is blinking when the unit is
switched on
Handpiece error
Air pressure error
Pressure not adjustable
Pressure reduction during operation
Software error
50
Error code
Solutions
--
Check the different electrical connections
Check the mains electricity supply
Check/replace the fuses and replace them with the type indicated on the back of the unit
Send the unit to your authorized EMS repair center
--
Send the unit to your authorized EMS repair center
1 LED
Check the different connections of the unit
Check the connection of the handpiece
Test the unit with another handpiece
2 LEDs
Check the different air connections
Check the air pressure supply is within the values indicated on the back of the unit
Check the compressor
Check the connection between the handpiece and unit
Test the unit with another handpiece
Send the unit to your authorized EMS repair center
--
Switch OFF/ON the unit
Check the different air connections
Check the air pressure supply is within the values indicated on the back of the unit
Check the compressor
Send the unit to your authorized EMS repair center
Repetitive
Check the different air connections
"beep"
Check the compressed air supply
Contact your authorized EMS repair center
3 LEDs
Send the unit to your authorized EMS repair center