2.1
Dangers and Risks
Technique and procedures
Only the physician can evaluate the clinical factors involved with each patient
and determine if the use of this device is indicated. The physician must
determine the specific technique and procedure that will accomplish the desired
clinical effect.
Factory settings
Check all factory settings. Factory settings are not mandatory settings for the
physician. The physician is responsible for all settings affecting the surgical
procedure.
Original accessories
For your own safety and that of your patient, use only original accessories.
Not explosion-proof
The device is not explosion-proof. Do not use in an area where flammable
anesthetic gases are present.
Risk of electrical shock
To prevent electrical shock, do not open this device. Never open this device
yourself. Refer servicing to qualified service personnel.
Professional qualification
This manual does not include descriptions or instructions for surgical
procedures/techniques. It is also not intended to introduce to or train physicians
in the use of surgical techniques. Medical accessories and devices may be used
only by physicians and medical assistants under the direction of a physician with
the appropriate technical/medical qualification.
Function test
The function test must be performed prior to each operation.
Sterile mediums and accessories
Always work exclusively with sterile substances and mediums, sterile fluids and
sterile accessories.
Disposal of irrigation fluid
Comply with local hygiene rules and regulations when disposing off the
irrigation fluid.
Safety Notes
DANGER
DANGER
DANGER
DANGER
DANGER
DANGER
DANGER
DANGER
DANGER
3
EN