LEVEL 4
■ transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Select
the function and press the START key.
The carriage moves to the transport
position. In the top left-hand corner of
the display you will see
refer to chapter 6 Transport).
■ synchronized/non-synchronized
mode
Motors A and B can run in synchro-
nized or non-synchronized mode.
synchronized:
Motors A and B perform a synchro-
nized movement according to the
ankle joint's physiological motion pat-
tern:
From the middle position the carriage
first goes to the maximum eversion
value. Afterwards, the carriage will si-
multaneously move to the maximum
inversion value and to the maximum
plantar flexion value, and then simul-
taneously to the maximum dorsal ex-
tension value and to the value halfway
between inversion and eversion. After
reaching this position, the cycle
restarts: maximum eversion at the
same time as halfway between dorsal
extension / plantar flexion, etc.
During operation the synchronized
mode is indicated in the top left-hand
corner by the symbol
non-synchronized:
Both motors run independently of
each other within the selected ranges
of motion.
To enable the non-synchronized
mode, disable the synchronized
mode.
During operation the non-synchro-
nized mode is indicated in the top
left-hand corner by the symbol
default setting:
synchronized mode enabled
Patient hazard –
The synchronized operation is
recommended as a general rule.
However, medical and/or therapeutic
indications may call for the non-syn-
chronized mode. In the non-synchro-
nized mode, particular caution and
attention on the part of the physi-
(also
cian/therapist is required to rule out
any risk for the patient.
■ total therapy time
ARTROMOT
chip card
The total therapy time is the added
sum of operating hours.
If the device is used by only one pa-
tient, this time is equivalent to the du-
ration of all the patient's therapy
sessions.
Under menu item "total therapy time"
of ARTROMOT
chip card you can view each pa-
tient's total therapy time (duration of
all the patient's therapy sessions).
Deleting the stored therapy time
Press and hold the parameter key for
5 seconds or select the New Patient
function.
■ Service MENU
For service purposes only, refer to
Service Manual.
.
Reminder:
You save the set parameter values by
pressing the STOP key.
.
59
Caution
®
-SP3 without patient
®
-SP3 models with