and the carriage moves in the oppo-
site direction.
This stretch cycle is repeated ten
times.
After that the carriage moves to the
programmed maximum eversion
value and restarts the inversion
stretch cycle.
default setting: disabled
Note!
- If a pause is programmed when you
enable the function, the carriage
will stop for the duration of the
pause each time the maximum
stretching value is attained.
- Enabling the stretching in inversion
function automatically disables
motor A. The only movement per-
formed is inversion / eversion.
- It is not possible to combine a sec-
ond stretching function with the
special function "stretching in inver-
sion".
- During operation the display will
show the symbol
■ stretching in eversion
With the special "stretching in ever-
sion" function the carriage gently
stretches the joint towards the out-
side.
Starting at the middle position the
carriage will first move to the pro-
grammed inversion value and then to
the programmed eversion value.
Subsequently the carriage reverses 5°
towards inversion and then moves
very slowly back again to the pro-
grammed eversion value (display
for the right foot and
left foot). After that it attempts to at-
tain another 5°, moving even slower
than before (display
foot and
for the left foot).
If a high resistance toward the addi-
tional 5° is sensed, the reverse on
load function is automatically acti-
vated and the carriage moves in the
opposite direction.
.
for the
for the right
64
This stretch cycle is repeated ten
times.
After that the carriage moves to the
programmed maximum inversion
value and restarts the eversion
stretch cycle.
default setting: disabled
Note!
- If a pause is programmed when you
enable the function, the carriage
will stop for the duration of the
pause each time the maximum
stretching value is attained.
- Enabling the stretching in eversion
function automatically disables
motor A. The only movement per-
formed is inversion / eversion.
- It is not possible to combine a sec-
ond stretching function with the
special function "stretching in ever-
sion".
- During operation the display will
show the symbol
Reminder:
You save the selected parameter values
by pressing the STOP key.
.