ARTROMOT SP3 Instrucciones Para El Uso página 61

Ocultar thumbs Ver también para SP3:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 112
Note!
LEVEL 1: equivalent to level 1 of
the Standard model (see: 5.3)
LEVEL 2: equivalent to level 2 of
the Standard model (see: 5.3)
LEVEL 3: equivalent to level 3 of
the Standard model (see: 5.3)
LEVEL 6: equivalent to level 4 of
the Standard model (see: 5.3)
LEVEL 4
■ warm-up protocol
During warm-up, the patient will
slowly become used to the set maxi-
mum dorsal extension / plantar flex-
ion and inversion / eversion values,
starting from the middle position.
The warm-up protocol starts in the
middle between the two maximum
values set for dorsal extension / plan-
tar flexion and inversion / eversion.
In the beginning, the range of inver-
sion / eversion motion increases with
each cycle, until the programmed
maximum values are reached after a
total of 15 cycles. In this phase,
motor A (dorsal extension / plantar
flexion) is inactive.
Then the carriage stops again in the
middle position and starts increasing
the range of dorsal extension / plantar
flexion motion with each cycle, until
the programmed maximum values are
reached after a total of 15 cycles. In
this phase, motor B (inversion / ever-
sion) is inactive.
After warm-up the device enters the
normal mode.
During warm-up, the display will show
the symbol
in the upper, left-
hand corner.
default setting: disabled
■ isolation program
For this special function, both motors
are turned on, but they will never
move simultaneously.
These are the steps of the special
function:
- At first motor B moves to the pro-
grammed maximum values for
inversion and eversion for ten
cycles, then it stops. In this phase,
motor A is disabled (symbol for
motor A:
)
- Then motor A will move to the pro-
grammed maximum values for dor-
sal extension and plantar flexion for
ten cycles and stops. The full cycle
will start over with motor B. Motor B
is disabled while motor A is operat-
ing (symbol for motor B:
- The stop position at the end of each
tenth cycle (for both dorsal exten-
sion / plantar flexion and inversion /
eversion) can be selected in steps of
25% between 0% and 100% of the
programmed maximum range of
motion.
- Steps 1 and 2 can be repeated as
often as needed. You can stop treat-
ment with the STOP key; after the
programmed therapy duration, it will
stop automatically.
During warm-up, the display will show
the symbol
corner.
default setting: disabled
Follow these steps to program the
special function:
- Access menu level 4, then press the
parameter key to select the special
function.
- The display will show the symbol for
the special function, its status (circle
with or without check mark) as well
as two boxes with the selected per-
centages for motor A and motor B
stop positions.
The circle appears in a box.
- Activate the function by pressing the
"+" key (plus). Deactivate the func-
tion by pressing the "-" key (minus).
In this case it is not possible to acti-
vate/deactive the function by press-
ing the parameter key again.
61
)
in the upper, left-hand

Publicidad

Tabla de contenido
loading

Productos relacionados para ARTROMOT SP3