8. Specifications
Input ratings:
100 – 240 V AC /
50 – 60 Hz
Current
850 - 370 mA
consumption:
Fuses:
2 x T1A (slow-blow)
Protection class: I
Applied part:
type B
Max. load on
20 kg
carriage:
Dimensions (transport position):
Length:
78.0 cm
Width:
42.0 cm
Height:
39.5 cm
Weight:
11 kg
Materials used: ABS, POM (Delrin
100), PUR, PA, FR4,
aluminum, stainless
steel, brass
MDD:
class 2a
Standards
compliance:
IEC 60601-1:1988
+ A1:1991
+ A2: 1995
Certification:
ANSI / UL 60601-1
CAN / CSA C22.2
No. 601.1
EMC
IEC 60601-1-2:2001
(electromagnetic compatibility)
Ambient conditions (storage, transport)
Temperature:
Relative humidity: 20% to 85%
Atmospheric
pressure:
Ambient conditions (operation)
Temperature:
Relative humidity: 30% to 75%
Atmospheric
pressure:
____________________________________
Subject to change without notice
(01/07)
68
-24 ºC to +60 ºC
(-12 to +140 °F)
700 hPa to 1060 hPa
+10 ºC to +40 ºC
(50 to +104 °F)
700 hPa to 1060 hPa