Environmental Protection Statement; Specifications - ARTROMOT SP3 Instrucciones Para El Uso

Ocultar thumbs Ver también para SP3:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 112
7. Environmental
Protection Statement
The product described in this operation
manual must not be disposed of with
unsorted household or municipal waste.
It requires separate disposal. Please
contact ORMED for information about
the possible recycling of the product.

8. Specifications

Input ratings:
100 – 240 V AC /
50 – 60 Hz
Current
850 - 370 mA
consumption:
Fuses:
2 x T1A (slow-blow)
Protection class: I
Applied part:
type B
Max. load on
20 kg
carriage:
Dimensions (transport position):
Length:
78.0 cm
Width:
42.0 cm
Height:
39.5 cm
Weight:
11 kg
Materials used: ABS, POM (Delrin
100), PUR, PA, FR4,
aluminum, stainless
steel, brass
MDD:
class 2a
Standards
compliance:
IEC 60601-1:1988
+ A1:1991
+ A2: 1995
Certification:
ANSI / UL 60601-1
CAN / CSA C22.2
No. 601.1
EMC
IEC 60601-1-2:2001
(electromagnetic compatibility)
Ambient conditions (storage, transport)
Temperature:
Relative humidity: 20% to 85%
Atmospheric
pressure:
Ambient conditions (operation)
Temperature:
Relative humidity: 30% to 75%
Atmospheric
pressure:
____________________________________
Subject to change without notice
(01/07)
68
-24 ºC to +60 ºC
(-12 to +140 °F)
700 hPa to 1060 hPa
+10 ºC to +40 ºC
(50 to +104 °F)
700 hPa to 1060 hPa

Publicidad

Tabla de contenido
loading

Productos relacionados para ARTROMOT SP3

Tabla de contenido